 +====== conclusion ======
 +
 +pro:
 +  * fulfils all initial power supply criteria 
 +  * **set output is within accuracy**
 +  * good design, lay out, quality components
 +  * fast first contact response
 +
 +contra:
 +  * output is not switched of completely
 +  * **measured value is up to 0.4 V lower than output voltage**
 +  * second digit after . of voltage is always 0 and not settable (it should not be displayed in this case)
 +  * second digit after . of voltage is always 0 and does not show measured value (the actual measured data is precise enough to show this digit)
 +  * can skip fast knob turns
 +  * USB cable does no fit in port because of the indent
 +  * information in english programming manual missing
 +  * small mistakes in programming manual
 +  * can not be calibrated by end user, and vendor only calibrates if under warranty
 +  * no contact support